• DocumentCode
    832927
  • Title

    The dependence of the energy resolution of gas proportional scintillation counters on the scintillation region to photomultiplier distance

  • Author

    Dos Santos, J.M.F. ; Bento, A.C.S.S.M. ; Conde, C.A.N.

  • Author_Institution
    Dept. of Phys., Coimbra Univ., Portugal
  • Volume
    39
  • Issue
    4
  • fYear
    1992
  • fDate
    8/1/1992 12:00:00 AM
  • Firstpage
    541
  • Lastpage
    545
  • Abstract
    The energy resolution for broad X-ray beams of gas proportional scintillation counters with medium to large size windows, parallel grids and no focusing system is discussed. The variation of the pulse amplitude with the radial distance is calculated for several distances d taken from the bottom of the scintillation region to the photomultiplier´s photocathode (in the 0 to 50 mm range), and compared with experimental results. It is shown that for a two-inch photomultiplier, d distances from 10 to 20 mm, commonly used in the detector designs, represent a worst-case situation
  • Keywords
    photomultipliers; scintillation counters; broad X-ray beams; energy resolution; focusing system; gas proportional scintillation counters; parallel grids; photomultiplier distance; photomultiplier´s photocathode; pulse amplitude; radial distance; two-inch photomultiplier; windows; Cathodes; Electrons; Energy resolution; Photomultipliers; Radiation detectors; Scintillation counters; Solid scintillation detectors; Voltage; X-ray detection; X-ray detectors;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/23.159662
  • Filename
    159662