Title :
The dependence of the energy resolution of gas proportional scintillation counters on the scintillation region to photomultiplier distance
Author :
Dos Santos, J.M.F. ; Bento, A.C.S.S.M. ; Conde, C.A.N.
Author_Institution :
Dept. of Phys., Coimbra Univ., Portugal
fDate :
8/1/1992 12:00:00 AM
Abstract :
The energy resolution for broad X-ray beams of gas proportional scintillation counters with medium to large size windows, parallel grids and no focusing system is discussed. The variation of the pulse amplitude with the radial distance is calculated for several distances d taken from the bottom of the scintillation region to the photomultiplier´s photocathode (in the 0 to 50 mm range), and compared with experimental results. It is shown that for a two-inch photomultiplier, d distances from 10 to 20 mm, commonly used in the detector designs, represent a worst-case situation
Keywords :
photomultipliers; scintillation counters; broad X-ray beams; energy resolution; focusing system; gas proportional scintillation counters; parallel grids; photomultiplier distance; photomultiplier´s photocathode; pulse amplitude; radial distance; two-inch photomultiplier; windows; Cathodes; Electrons; Energy resolution; Photomultipliers; Radiation detectors; Scintillation counters; Solid scintillation detectors; Voltage; X-ray detection; X-ray detectors;
Journal_Title :
Nuclear Science, IEEE Transactions on