• DocumentCode
    833361
  • Title

    Bulk Performance of Improved Mercuric Iodide Nuclear Detectors

  • Author

    Schieber, M. ; Beinglass, I. ; Dishon, G. ; Holzer, A. ; Yaron, G.

  • Author_Institution
    School of Applied Science and Technology The Hebrew University of Jerusalem, Jerusalem, Israel
  • Volume
    25
  • Issue
    1
  • fYear
    1978
  • Firstpage
    644
  • Lastpage
    648
  • Abstract
    The bulk homogeneity of HgI2 detectors was evaluated by the nuclear spectral response of several detectors fabricated from the same iodine treated crystal grown by TOM (Temperature Oscillating Method) in a horizontal furnace. The 241 Am and 137CS spectral resolution is shown as a function of the detector thickness which ranges from 50 to 1000 ¿m and of the bias voltage. The lowest ¿T value for holes measured with 55Fe on these detectors is about 10-6 cm2V-l. Nevertheless, the spectral resolution for 137Cs indicates that these detectors must have spots with a higher value of ¿Th Uniformity studies by scanning one of the detectors with 241Am are also given and discussed. In addition a 60Co spectrum taken with a 1 mm thick HgI2 detector is shown for the first time.
  • Keywords
    Charge carrier processes; Crystalline materials; Crystals; Electron traps; Furnaces; Gas detectors; Phase detection; Temperature; Thickness measurement; Voltage;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1978.4329386
  • Filename
    4329386