• DocumentCode
    833403
  • Title

    Pressurized Xenon-Filled Multiwire Proportional Chamber for Radionuclide Imaging

  • Author

    Bolon, Craig ; Deutsch, Martin ; Lanza, Richard ; Burns, Ronald ; Judy, Philip ; Zimmerman, Robert

  • Author_Institution
    Laboratory for Nuclear Science, Massachusetts Institute of Technology, Cambridge, Massachusetts 02139
  • Volume
    25
  • Issue
    1
  • fYear
    1978
  • Firstpage
    661
  • Lastpage
    664
  • Abstract
    A pressurized multiwire proportional chamber radionuclide imaging system has been constructed which has operated at pressures of up to 11 atmospheres absolute using a xenon-methane gas mix. The experimental detector has an active imaging area of 25 sq. cm. and an effective depth of 11 cm. An all digital readout system has been constructed with an intrinsic resolution of 1.3 mm FWHM. The FWHM has been measured for various radionuclides: iodine-125 (28 keV), 1.3 mm; thallium - 201 (75 keV), 2 mm; technetium-99 m (140 keV), 3 mm. Our primary goal has been the investigation of properties of such detectors that offer advantages over conventional gamma cameras, including high data rate and improved spatial resolution, particularly for isotopes with photon energies below 100 keV. We have fabricated a gas purification system which permits long term operation of the system without degradation and have investigated construction techniques to produce reliable devices which could be used in a hospital environment.
  • Keywords
    Atmosphere; Atmospheric measurements; Cameras; Degradation; Gamma ray detection; Gamma ray detectors; Hospitals; Isotopes; Purification; Spatial resolution;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1978.4329390
  • Filename
    4329390