Title :
A 400 Anode Chevron Microchannel Plate PMT for High Energy Application
Author :
Oba, Koichiro ; Sugiyama, Masaru ; Suzuki, Yoshiji ; Yoshimura, Yoshio
Author_Institution :
Hamamatsu TV Co., Ltd., Ichino-cho 1126, Hamamatsu, Japan
Abstract :
A proto type multianode photomultiplier tube incorporating a chevron type MCPs electron multiplier and 400 multi anodes was developed for high speed image pick-up at low light levels such as in detection of rings of Cherenkov light. The diameter of and mean spacing of each anode are 0.3mm and 0.65mm respectively in a rectangular array of 16mmÃ12mm. The MCP used in the tube were fabricated at HTV and had 24.9mm diameter, 20mm useful diameter, 15¿m pore diameter and 45 length to pore diameter ratio. Electron multiplication and dark current for each MCP were about 104 and 1pA at 1000v respectively. 0.6mm spatial resolution over anode surface and about 600psec rise time of output signal were obtained.
Keywords :
Anodes; Cathodes; Dark current; Electrons; Event detection; Image sensors; Image storage; Microchannel; Sensor arrays; Spatial resolution;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1979.4329656