DocumentCode :
835998
Title :
Improvement of the Light Collection in Scintillation Detectors
Author :
Klein, Horst ; Scholermann, Horst
Volume :
26
Issue :
1
fYear :
1979
Firstpage :
373
Lastpage :
377
Abstract :
The influence of geometrical and optical parameters on the light collection within scintillator assemblies has been investigated. Monte-Carlo calculations predicted a strong dependence on the design of the light guide and the effective reflectivity of its covering. Excellent agreement of experimental and theoretical data is shown for a small " standard"-scintillator (5.08 cm x 5.08 cm in diameter) as well as for largearea systems (5.08 cm x 25.4 cm in diameter). In both cases partially coated light guides should be used to restrict the local dependence of the light transmission to less than 2 % or 4 % resp. over the total volume. The averaged absolute transmission is calculated to be approximately 80 % or 45 % resp. for these optimized setups.
Keywords :
Assembly; Detectors; Energy resolution; Geometrical optics; Neutrons; Optical refraction; Photomultipliers; Production; Reflectivity; Scintillation counters;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1979.4329660
Filename :
4329660
Link To Document :
بازگشت