• DocumentCode
    837158
  • Title

    Performance of the ultra wide bore 900 MHz NMR magnet at the national high magnetic field laboratory

  • Author

    Dixon, Iain R. ; Markiewicz, W. Denis ; Brey, William W. ; Shetty, Kiran K.

  • Author_Institution
    Nat. High Magnetic Field Lab., Tallahassee, FL, USA
  • Volume
    15
  • Issue
    2
  • fYear
    2005
  • fDate
    6/1/2005 12:00:00 AM
  • Firstpage
    1334
  • Lastpage
    1337
  • Abstract
    A 900 MHz NMR magnet with a room temperature, clear bore of 105 mm was recently brought to full field at the NHMFL. The magnet was ramped to 21.1 T on its first run without incurring a quench, since installation into its final cryostat was completed. Issues on superconductor stability were present during magnet charging and are discussed. In addition, the magnet performance in terms of field uniformity and of field stability, with the implementation of a novel design concept of current injection, is presented.
  • Keywords
    nuclear magnetic resonance; quenching (thermal); stability; superconducting magnets; 105 mm; 21.1 T; 900 MHz; National High Magnetic Field Laboratory; current injection; field stability; field uniformity; magnet charging; magnet performance; superconductor stability; ultra wide bore NMR magnet; Boring; High temperature superconductors; Laboratories; Magnetic fields; Niobium compounds; Nuclear magnetic resonance; Stability; Superconducting coils; Superconducting magnets; Titanium compounds; NMR magnets; stability; superconductivity; uniformity;
  • fLanguage
    English
  • Journal_Title
    Applied Superconductivity, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1051-8223
  • Type

    jour

  • DOI
    10.1109/TASC.2005.849588
  • Filename
    1439888