DocumentCode :
838040
Title :
The Use of Compact Cyclotrons for Producing Fast Neutrons for Therapy in a Rotatable Isocentric Gantry
Author :
Tom, James L.
Author_Institution :
The Cyclotron Corporation Berkeley, California 94710
Volume :
26
Issue :
2
fYear :
1979
fDate :
4/1/1979 12:00:00 AM
Firstpage :
2294
Lastpage :
2298
Abstract :
This paper discusses the design and performance of a fast neutron therapy system which transports a cyclotron beam onto a beryllium target mounted in a rotatable isocentric gantry structure. Fast neutrons produced by either the proton or deuteron on beryllium reaction are directed towards the isocenter with appropriate shielding, collimation and dosimetry to control and monitor the radiation delivered onto a defined patient area. The beam optics and shielding considerations of such a radiotherapy apparatus are discussed. This paper also presents the dosimetric properties of neutron beams from different proton energies on beryllium and lithium targets.
Keywords :
Cyclotrons; Dosimetry; Medical treatment; Neutrons; Optical beams; Optical collimators; Particle beams; Patient monitoring; Protons; Radiation monitoring;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1979.4329861
Filename :
4329861
Link To Document :