An improved multicrystal 2-D BGO detector for PET
Author
Rogers, J.G. ; Taylor, A.J. ; Rahimi, M.F. ; Nutt, R. ; Andreaco, M. ; Williams, C.W.
Author_Institution
TRIUMF, Vancouver, BC, Canada
Volume
39
Issue
4
fYear
1992
fDate
8/1/1992 12:00:00 AM
Firstpage
1063
Lastpage
1068
Abstract
The authors evaluate and compare two 2-D array detectors for positron emission tomography (PET). Both consisted of an 8×8 array of small BGO crystals coupled to a 2×2 array of photomultiplier tubes. The depth of the crystals was 3 cm in one detector and 2 cm in the other. The 2-cm detector is superior in terms of material costs, in energy resolution per crystal, and in its ability to clearly identify the crystal containing the primary interaction. A flexible and robust algorithm is presented for crystal identification in such array detectors. The prospect of obtaining still better spatial resolution from such block detectors, with increased numbers of crystals, is discussed
Keywords
beryllium compounds; biomedical equipment; computerised tomography; electron detection and measurement; radioisotope scanning and imaging; scintillation counters; BGO crystals; Be4Ge3O12; PET; array detectors; crystal identification; photomultiplier tubes; positron emission tomography; spatial resolution; Costs; Crystalline materials; Crystals; Detectors; Energy resolution; Photomultipliers; Positron emission tomography; Robustness; Sensor arrays; Spatial resolution;