  • Abstract
    Incoherent averaging of measurements made at different look directions can be used to reduce speckle noise in medical ultrasound images. An analytic expression for the correlation between two measurements made at different spatial positions is derived. Using this expression, the optimum aperture displacement for efficient incoherent averaging (i.e. compounding) is computed and found to equal approximately one-half the aperture length.<>
  • Keywords
    acoustic imaging; biomedical ultrasonics; acoustic imaging; compound image generation; medical ultrasound; optimum aperture displacement; speckle noise; Apertures; Biomedical imaging; Image generation; Length measurement; Noise measurement; Noise reduction; Position measurement; Speckle; Ultrasonic imaging; Ultrasonic variables measurement;