• DocumentCode
    842396
  • Title

    A 128 × 128 CMOS biosensor array for extracellular recording of neural activity

  • Author

    Eversmann, Björn ; Jenkner, Martin ; Hofmann, Franz ; Paulus, Christian ; Brederlow, Ralf ; Holzapfl, Birgit ; Fromherz, Peter ; Merz, Matthias ; Brenner, Markus ; Schreiter, Matthias ; Gabl, Reinhard ; Plehnert, Kurt ; Steinhauser, Michael ; Eckstein, Ge

  • Author_Institution
    Corporate Res., Infineon Technol. AG, Munich, Germany
  • Volume
    38
  • Issue
    12
  • fYear
    2003
  • Firstpage
    2306
  • Lastpage
    2317
  • Abstract
    Sensor arrays are a key tool in the field of neuroscience for noninvasive recording of the activity of biological networks, such as dissociated neurons or neural tissue. A high-density sensor array complementary metal-oxide-semiconductor chip is presented with 16 K pixels, a frame rate of 2 kiloframes per second, and a pitch of 7.8 μm × 7.8 μm for imaging of neural activity. The related circuit and system issues as well as process aspects are discussed. A mismatch-canceling calibration circuitry with current mode signal representation is used. Results from first biological experiments are presented, which prove full functionality of the chip.
  • Keywords
    CMOS integrated circuits; bioelectric potentials; biomedical electrodes; biomedical electronics; biomedical transducers; biosensors; neurophysiology; CMOS biosensor array; bioelectric potentials; biomedical transducers; current mode signal representation; extracellular recording; high-density sensor array; mismatch-canceling calibration circuitry; multi-electrode array; neural activity; noninvasive recording; pixel circuit; Biological tissues; Biosensors; Calibration; Circuits and systems; Extracellular; Image sensors; Neurons; Neuroscience; Pixel; Sensor arrays;
  • fLanguage
    English
  • Journal_Title
    Solid-State Circuits, IEEE Journal of
  • Publisher
    ieee
  • ISSN
    0018-9200
  • Type

    jour

  • DOI
    10.1109/JSSC.2003.819174
  • Filename
    1253878