DocumentCode
842881
Title
Long-term stimulation and recording with a penetrating microelectrode array in cat sciatic nerve
Author
Branner, Almut ; Stein, Richard B. ; Fernandez, Eduardo ; Aoyagi, Yoichiro ; Normann, Richard A.
Author_Institution
Dept. of Bioeng., Univ. of Utah, Salt Lake City, UT, USA
Volume
51
Issue
1
fYear
2004
Firstpage
146
Lastpage
157
Abstract
We studied the consequences of long-term implantation of a penetrating microelectrode array in peripheral nerve over the time course of 4-6 mo. Electrode arrays without lead wires were implanted to test the ability of different containment systems to protect the array and nerve during contractions of surrounding muscles. Treadmill walking was monitored and the animals showed no functional deficits as a result of implantation. In a different set of experiments, electrodes with lead wires were implanted for up to 7 mo and the animals were tested at 2-4 week intervals at which time stimulation thresholds and recorded sensory activity were monitored for every electrode. It was shown that surgical technique highly affected the long-term stimulation results. Results between measurement sessions were compared, and in the best case, the stimulation properties stabilized in 80% of the electrodes over the course of the experiment (162 days). The recorded sensory signals, however, were not stable over time. A histological analysis performed on all implanted tissues indicated that the morphology and fiber density of the nerve around the electrodes were normal.
Keywords
bioelectric phenomena; microelectrodes; neuromuscular stimulation; prosthetics; 162 day; 2 to 4 week; 4 to 6 month; Pb; cat sciatic nerve; functional neuromuscular stimulation; histological analysis; implanted tissues; lead wires; long-term implantation; long-term stimulation; muscle contractions; nerve fiber density; nerve morphology; neuroprosthesis; penetrating microelectrode array; peripheral nerve; recording; sensory activity; treadmill walking; Animals; Electrodes; Legged locomotion; Microelectrodes; Monitoring; Muscles; Protection; Surgery; System testing; Wires; Action Potentials; Animals; Cats; Electric Stimulation; Electromyography; Equipment Failure Analysis; Locomotion; Microelectrodes; Monitoring, Ambulatory; Muscle, Skeletal; Prosthesis Implantation; Sciatic Nerve;
fLanguage
English
Journal_Title
Biomedical Engineering, IEEE Transactions on
Publisher
ieee
ISSN
0018-9294
Type
jour
DOI
10.1109/TBME.2003.820321
Filename
1254004
Link To Document