• DocumentCode
    844970
  • Title

    "Positology"-the search for suitable detector arrangements for a positron ECT with continuous rotation

  • Author

    Tanaka, Eiichi ; Nohara, Norimasa ; Yamamoto, Mikio ; Tomitani, Takehiro ; Murayama, Hideo ; Ishimatsu, Kenji ; Takami, Katsumi

  • Author_Institution
    National Institute of Radiological Sciences, Anagawa, Chiba-shi, Japan
  • Volume
    26
  • Issue
    2
  • fYear
    1979
  • fDate
    4/1/1979 12:00:00 AM
  • Firstpage
    2728
  • Lastpage
    2731
  • Abstract
    This paper proposes a continuously rotating positron ECT system, in which the detectors are arranged on a circular ring with "non-uniform" spacing so as to provide much finer sampling interval in the projections than the conventional circular ring system. The address signals on the coincidence events are read out through a rotary photo-coupler, and the detector system can be rotated at a desired speed. Suitable detector arrangements are searched which provide sufficiently fine sampling interval with reasonable sampling density uniformity. Assuming about 50 detectors are arranged on a circle, the sampling characteristics of various detector arrangements have been examined. An iterative search method has also been applied to find a reasonable arrangement. The best arrangement has been obtained by the iterative method. It has an effective sampling interval of the order of (1/200)R, where R is the radius of the detector array, and the sampling density uniformity after 5-point smoothing is -7.8 ˜/spl I.chemo/18 % in the range of /spl I.chemo/ R/2.
  • Keywords
    Biomedical imaging; Detectors; Electrical capacitance tomography; Event detection; Image sampling; Iterative methods; Motion detection; Positrons; Sampling methods; Sensor arrays;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1979.4330524
  • Filename
    4330524