  • Abstract
    The paper deals with selected topics and problem areas of x-ray tubes, particularly those with rotating anodes, as they relate to CT scanner requirements. Only the main limiting factors will be treated, such as "rotor ripple", heat problems, off-focus x-rays, and high voltage stability. Some basic considerations are given to super large x-ray source assemblies for ultra fast scanners. Because of the extent of x-ray tube physics and engineering, the discussion had to be kept short.
  • Keywords
    Anodes; Apertures; Collimators; Computed tomography; Fluctuations; Image reconstruction; Physics; Rotors; Voltage; X-ray imaging;