DocumentCode
845197
Title
Physics of X-Ray Tubes for CT Scanners
Author
Braun, Martin
Author_Institution
Radiologic Sciences Inc. (Pfizer), Santa Clara, CA
Volume
26
Issue
2
fYear
1979
fDate
4/1/1979 12:00:00 AM
Firstpage
2840
Lastpage
2844
Abstract
The paper deals with selected topics and problem areas of x-ray tubes, particularly those with rotating anodes, as they relate to CT scanner requirements. Only the main limiting factors will be treated, such as "rotor ripple", heat problems, off-focus x-rays, and high voltage stability. Some basic considerations are given to super large x-ray source assemblies for ultra fast scanners. Because of the extent of x-ray tube physics and engineering, the discussion had to be kept short.