Abstract :
The accuracy of measurements of the quantized Hall resistivity (QHR) by a direct-current-comparator (DCC)-type potentiometer is estimated based on a linearity calibration of the potentiometer using a Josephson potentiometer. The power coefficient contribution to the nonlinearity of the DCC potentiometer was found to be 0.15±0.02 p.p.m./(100 mV)2 at 18.5±0.5°C in 1985 and 0.21±0.03 p.p.m./(100 mV)2 at 20.5±0.5°C in 1988. The possibility of accurate measurements of the ratio QHR/R STD with uncertainties less than 0.05 p.p.m. by the DCC potentiometer is discussed
Keywords :
calibration; constants; electric resistance measurement; potentiometers; quantum Hall effect; Josephson potentiometer; calibration; constants; direct current comparator; nonlinearity; power coefficient; quantized Hall resistivity; Calibration; Conductivity; Current measurement; Electrical resistance measurement; Frequency; Josephson junctions; Linearity; Measurement standards; Microwave antenna arrays; Potentiometers;