• DocumentCode
    845446
  • Title

    Josephson array voltage calibration system: operational use and verification

  • Author

    Steiner, Richard L. ; Field, Bruce F.

  • Author_Institution
    NBS, Gaithersburg, MD, USA
  • Volume
    38
  • Issue
    2
  • fYear
    1989
  • fDate
    4/1/1989 12:00:00 AM
  • Firstpage
    296
  • Lastpage
    301
  • Abstract
    A Josephson array system that maintains the US legal volt is described. This system is almost fully automated, operates with a typical precision of 0.009 μV, and readily allows US legal volt measurements weekly, or more frequently if desired. This system was compared to the previous volt-maintenance system, and agreement was found to within 0.03 p.p.m. This verification is limited by uncertainties in the resistive divider instruments of the previous system
  • Keywords
    Josephson effect; calibration; units (measurement); voltage measurement; Josephson array system; US legal volt measurements; resistive divider instruments; verification; volt-maintenance; voltage calibration; Calibration; Circuits; Frequency; Gunn devices; Instruments; Law; Legal factors; Millimeter wave measurements; Semiconductor device measurement; Voltage;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.192292
  • Filename
    192292