A 10-V Josephson voltage standard
Author
Hamilton, Clark A. ; Lloyd, Frances L. ; Chieh, Kao ; Goeke, Wayne C.
Author_Institution
NBS, Boulder, CO, USA
Volume
38
Issue
2
fYear
1989
fDate
4/1/1989 12:00:00 AM
Firstpage
314
Lastpage
316
Abstract
The design and operation of an 18992-Josephson-junction array, which can generate reference voltages up to 12 V, is described. This device has applications for the direct calibration of Zener reference standards, calibrators, and digital voltmeters at the 10-V level, and for very accurate linearity and ratio measurements. The modified design of the array resolves the problems of stability and power requirement
Keywords
Josephson effect; measurement standards; superconducting junction devices; voltage measurement; 10 V; Josephson voltage standard; Zener reference standards; calibration; digital voltmeters; linearity; measurement standard; power; ratio measurements; stability; Calibration; Critical current; Fabrication; Josephson junctions; Linearity; Measurement standards; Microstrip; Stability; Voltage; Voltmeters;