• DocumentCode
    845582
  • Title

    High accuracy calculable AC-DC transfer standards for the LF-30 MHz frequency range

  • Author

    Nomair, M. ; Harmans, Kees J.P.M.

  • Author_Institution
    Van Swinden Lab., Delft, Netherlands
  • Volume
    38
  • Issue
    2
  • fYear
    1989
  • fDate
    4/1/1989 12:00:00 AM
  • Firstpage
    342
  • Lastpage
    345
  • Abstract
    A model for a calculable AC-DC thermal transfer device is described. It is based on the transmission-line theory, augmented by corrections for skin effect and connection admittance. Experimental verification on four devices shows good to excellent agreement between the calculated and measured results. This enables the establishment of AC-DC transfer standards with total uncertainties ranging from less than 5 p.p.m. at 1 MHz to 100 p.p.m. at 30 MHz. Potential for extension to higher frequencies is discussed
  • Keywords
    electric current measurement; measurement standards; skin effect; transmission line theory; voltage measurement; 1 to 30 MHz; AC-DC transfer standards; connection admittance; model; skin effect; thermal transfer; transmission-line theory; Admittance; Connectors; Frequency; Impedance; Laboratories; Resistors; Tellurium; Transmission line theory; Voltage; Wire;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.192303
  • Filename
    192303