• DocumentCode
    845784
  • Title

    Precision AC measurements of temperature below 90 K

  • Author

    Levine, Paul D. ; Jaeger, K.B. ; Mellons, Brad E.

  • Author_Institution
    Lockheed Missiles & Spacecraft Co., Sunnyvale, CA, USA
  • Volume
    38
  • Issue
    2
  • fYear
    1989
  • fDate
    4/1/1989 12:00:00 AM
  • Firstpage
    428
  • Lastpage
    433
  • Abstract
    Specially designed sealed cells filled with ultrapure gas have been utilized for the realization of thermodynamic fixed points in the Primary Standards Laboratory. Accurate determination of the triple-point temperatures of Ar, N2, O2, Ne, and D2 was accomplished with a high-precision AC resistance bridge operating at a frequency of 30 Hz. Detailed data obtained for platinum resistance thermometers (PRTs) and rhodium-iron (Rh-Fe) sensors are presented. The levels of accuracy and precision achieved are evaluated
  • Keywords
    bridge instruments; critical points; electric sensing devices; iron alloys; low-temperature techniques; measurement standards; platinum; resistance thermometers; rhodium alloys; temperature measurement; 30 Hz; 90 K; AC measurements; Ar; D2; N2; Ne; O2; Primary Standards Laboratory; Pt resistance thermometer; Rh-Fe sensor; resistance bridge; sealed cells; thermodynamic fixed points; triple-point temperatures; ultrapure gas; Argon; Bridges; Frequency; Laboratories; Measurement standards; Platinum; Temperature measurement; Temperature sensors; Thermal resistance; Thermodynamics;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/19.192320
  • Filename
    192320