Reduction of Polarization in Mercuric Iodide Nuclear Radiation Detectors
Author :
Holzer, A. ; Schieber, M.
Author_Institution :
School of Applied Science and Technology, Hebrew University of Jerusalem Jerusalem, Israel
Volume :
27
Issue :
1
fYear :
1980
Firstpage :
266
Lastpage :
271
Abstract :
The majority of HgI2 detectors which have an energy resolution less than 5% for 241Am do not have severe plarization effects at all. Polarization effects are reported for detectors having energy resolution greater than 5% for 241Am in an attempt to discover if polarization is mainly due to either a nonoptimal fabrication or to severe trapping in the bulk crystalline material. An improved fabrication procedure using a rapid etch and evaporated palladium contacts is reported. This procedure has produced polarization free detectors which were refabricated from detectors which previously showed polarization.
Keywords :
Crystalline materials; Crystals; Degradation; Energy resolution; Fabrication; Leak detection; Leakage current; Polarization; Radiation detectors; Voltage;