Abstract :
A new centroid finding method for all detectors where the signal charge is collected or induced on strips or wires, or on subdivided resistive electrodes, is presented. The centroid of charge is determined by convolution of the sequentially switched outputs from these subdivisions or from the strips with a linear centroid finding filter. The position line width is inversely proportional to N3/2, where N is the number of subdivisions.
Keywords :
Capacitance; Delay lines; Electrodes; Laboratories; Position sensitive particle detectors; Quantization; Signal detection; Signal resolution; Strips; Wires;