DocumentCode :
848676
Title :
Design Criteria for Nulti-Slice Positron Emission-Computed Tomography Detector Systems
Author :
Carroll, L.R. ; Hendry, G.O. ; Currin, J.D.
Author_Institution :
The Cyclotron Corporation Berkeley, California
Volume :
27
Issue :
1
fYear :
1980
Firstpage :
485
Lastpage :
488
Abstract :
Design criteria for single-slice positron emission computed tomography systems have been extensively discussed in the literature. We wish to extend the current methodology and suggest some additional tools for optimizing the design of multi-slice tomographic systems. Practical results using Bismuth Germanate coincidence detectors are shown. The effect of shielding design on overall system performance is evaluated with the help of a simple graphical technique.
Keywords :
Detectors; Equations; Event detection; Geometry; Imaging phantoms; Positrons; Scattering parameters; Space vector pulse width modulation; Timing; Tomography;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1980.4330873
Filename :
4330873
Link To Document :
بازگشت