DocumentCode :
848695
Title :
A Hybrid Electrode Array With Built-In Position Sensors for an Implantable MEMS-Based Cochlear Prosthesis
Author :
Wang, Jianbai ; Wise, Kensall D.
Author_Institution :
Dept. of Technol. Dev., Texas Instrum., Dallas, TX
Volume :
17
Issue :
5
fYear :
2008
Firstpage :
1187
Lastpage :
1194
Abstract :
A hybrid electrode array has been developed for use in an implantable MEMS-based cochlear prosthesis intended to restore hearing to the profoundly deaf. The array provides high-density stimulation and embedded position sensing to improve sound perception, minimize insertion damage, and optimize implant placement. Interfacing with a hermetically-packaged microcontroller and a bidirectional wireless interface over an eight-lead polymeric cable, the array integrates custom signal-processing electronics with thin-film lithographically defined IrO electrodes and polysilicon position sensors. The signal-processing chip (2.4 mm times 2.4 mm) operates from plusmn2.5 V, accepts 16-b commands, and performs command validation, stimulus current generation, site/sensor selection, offset compensation, and output-signal conditioning. The array contains nine polysilicon strain gauges to detect array position and tip contact. These strain sensors have gauge factors of 15-20, allowing the array tip position to be determined within 50 mum while providing tip-contact signals of more than 100 mV. The on-chip signal processing allows the use of a standard bus interface, reduces the parasitic capacitance and noise coupling on the connecting cables, allows the array shape to be imaged at a rate up to 10 times per second, and suppresses the output noise down to 3 mV.
Keywords :
bioMEMS; biomedical electrodes; prosthetics; silicon; MEMS-based cochlear prosthesis; Si; bidirectional wireless interface; built-in position sensors; command validation; eight-lead polymeric cable; gauge factors; hermetically-packaged microcontroller; high-density stimulation; hybrid electrode array; implant placement; insertion damage; offset compensation; output-signal conditioning; polysilicon position sensors; signal-processing chip; signal-processing electronics; site-sensor selection; sound perception; standard bus interface; stimulus current generation; strain sensors; tip contact detection; voltage 2.5 V to -2.5 V; Cochlear prosthesis; neural probes; piezoresistance; strain gauge;
fLanguage :
English
Journal_Title :
Microelectromechanical Systems, Journal of
Publisher :
ieee
ISSN :
1057-7157
Type :
jour
DOI :
10.1109/JMEMS.2008.928705