Title :
Positologica: A Positron ECT Device with a Continuously Rotating Detector Ring
Author :
Nohara, Norimasa ; Tanaka, Eiichi ; Tomitani, Takehiro ; Yamamoto, Mikio ; Murayama, Hideo ; Suda, Yoshio ; Endo, Masahiro ; Iinuma, Takeshi ; Tateno, Yukio ; Shishido, Fumio ; Ishimatsul, Kenji ; Ueda, Ken ; Takami, Katsumi
Author_Institution :
National Institute of Radiological Sciences, Anagawa, Chiba-shi 260, Japan
fDate :
6/1/1980 12:00:00 AM
Abstract :
A rotary positron emission computed tomography device is developed for human brain and animal studies. The device utilizes 64 rectangular BGO detectors arranged at unequal spacing on a circular ring. The detector ring is continuously rotated at a constant speed of 60 rpm or less. This single continuous motion of the ring with the detector array provides excellent sampling characteristics and high detector redundancy. The device has a field of view 24 cm in diameter with a slice thickness adjustable from 1 cm to 2 cm. Measured width of system response to a 2 mm diameter line source is 5.8 mm FWHM at the center and less than 9 mm FWHM within a circle 16 cm in diameter. Measured sensitivity including scattered coincidence events is about 17 kcps/pCi/ml for a 20 cm diameter water phantom and 2 cm thick slice.
Keywords :
Animals; Computed tomography; Detectors; Electrical capacitance tomography; Humans; Motion detection; Positrons; Radioactive decay; Sampling methods; Sensor arrays;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1980.4330985