• DocumentCode
    850179
  • Title

    Transient Humidity Measurements—Part I: Sensor Calibration and Characteristics

  • Author

    Wang, Yiheng Ian ; Simonson, Carey J. ; Besant, Robert W. ; Shang, Wei

  • Author_Institution
    Dept. of Mech. Eng., Saskatchewan Univ., Saskatoon, Sask.
  • Volume
    56
  • Issue
    3
  • fYear
    2007
  • fDate
    6/1/2007 12:00:00 AM
  • Firstpage
    1074
  • Lastpage
    1079
  • Abstract
    In this paper, a transient test facility is used to investigate the transient response of a polymer film electronic humidity sensor (capacitance type) following a step change in humidity and temperature. The response of the humidity sensor is measured for a step change in the humidity with and without a change in temperature, and the data are correlated with exponential functions with two time constants. The major time constant of the sensor is about 3 s for the isothermal test but increases by an order of magnitude when the sensor is exposed to a simultaneous change in temperature and humidity
  • Keywords
    calibration; electric sensing devices; humidity measurement; humidity sensors; transient response; humidity measurements; isothermal test; polymer film electronic humidity sensor; sensor calibration; transient response; transient test facility; Calibration; Capacitance; Capacitive sensors; Humidity measurement; Polymer films; Sensor phenomena and characterization; Temperature sensors; Test facilities; Time measurement; Transient response; Experiment; humidity; humidity sensor; response time; simultaneous heat and mass transfer; step change; temperature; transient calibration;
  • fLanguage
    English
  • Journal_Title
    Instrumentation and Measurement, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9456
  • Type

    jour

  • DOI
    10.1109/TIM.2007.894881
  • Filename
    4201025