• DocumentCode
    851219
  • Title

    Dependence of the performance of CsI-covered microstrip plate VUV photosensors on geometry: experimental results

  • Author

    Freitas, D.S.A.P. ; Veloso, J.F.C.A. ; dos Santos, J.M.F. ; Conde, C.A.N.

  • Author_Institution
    Dept. de Fisica, Coimbra Univ., Portugal
  • Volume
    49
  • Issue
    4
  • fYear
    2002
  • Firstpage
    1629
  • Lastpage
    1633
  • Abstract
    The behavior of a xenon-filled microstrip gas chamber instrumented with a microstrip plate with eight different microstrip geometries is reported. The best energy resolution achieved for 5.9-keV and for 22.1-keV X-rays was 13.6% and 7.0%, respectively. The same microstrip plate covered with a CsI film was used in place of the photomultiplier tube to detect the vacuum ultraviolet scintillation light produced in a driftless xenon-filled gas proportional scintillation counter. The relative amplitude, energy resolution,. optical positive feedback, and photoelectron extraction efficiency are discussed. The energy resolution obtained for 5.9-keV X-rays was 11.4%, which is better than that achieved with a standard proportional counter, about 13% for argon based proportional counters.
  • Keywords
    X-ray detection; caesium compounds; gas scintillation detectors; proportional counters; 22.1 keV; 5.9 keV; CsI; CsI-covered microstrip plate VUV photosensors; Xe; driftless xenon-filled gas proportional scintillation counter; energy resolution; optical positive feedback; photoelectron extraction efficiency; relative amplitude; vacuum ultraviolet scintillation light; xenon-filled microstrip gas chamber; Electron tubes; Energy resolution; Geometry; Instruments; Microstrip; Optical feedback; Optical films; Photomultipliers; Radiation detectors; X-rays;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2002.801698
  • Filename
    1043391