DocumentCode
851343
Title
Cardiac Imaging with a High Pressure Low Dead Time Multiwire Proportional Chamber
Author
Zimmerman, R.E. ; Holman, B.L. ; Fahey, F.H. ; Lanza, R.C. ; Cheng, C. ; Treves, S.
Author_Institution
Harvard Medical School, Joint Program in Nuclear Medicine, Brigham and Women´´s Hospital Medical Center, Boston, MA 02115
Volume
28
Issue
1
fYear
1981
Firstpage
55
Lastpage
56
Abstract
A sealed, 10 atmosphere MWPC has been constructed for cardiac imaging applications in nuclear medicine. Spatial resolution is < 3.5mm and count rate characteristics are a factor of 2 better than current cameras, resulting in improved statistical quality of information. Major applications are first pass radionuclide angiocardiograms with 9.3 m. Ta-178 and 4.9s Ir-191m. Animal and patient studies have been performed and the advantages of this system have been demonstrated.
Keywords
Anodes; Atmosphere; Biomedical imaging; Cameras; Cathodes; Detectors; Instruments; Isotopes; Nuclear medicine; Spatial resolution;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1981.4331139
Filename
4331139
Link To Document