• DocumentCode
    851863
  • Title

    Calibration of an alpha-beta moving filter particulates monitor

  • Author

    Klett, Alfred ; DeMey, Leo ; Erath, Wolfgang ; Nemecek, Peter

  • Author_Institution
    Berthold Technol. GmbH & Co. KG, Bad Wildbad, Germany
  • Volume
    49
  • Issue
    4
  • fYear
    2002
  • fDate
    8/1/2002 12:00:00 AM
  • Firstpage
    1932
  • Lastpage
    1934
  • Abstract
    Sampling and detection of airborne particulates is a dynamical process and depends on the detailed characteristics of monitoring instruments. A clear understanding of the impact of these details is essential for superior designs and optimized conditions of operation. Therefore, the new moving filter particulates monitor BAI 9128-ENV for environmental monitoring of airborne activity was carefully calibrated by injection of known activity concentrations. These experiments were conducted at the NRG calibration facilities in Petten, the Netherlands. The instrument´s calibration factor and its dynamic behavior were determined from these measurements. Different techniques for the compensation of natural activities were applied and investigated. Detection limits in the presence of radon were determined.
  • Keywords
    air pollution measurement; alpha-particle detection; beta-ray detection; calibration; radiation monitoring; silicon radiation detectors; BAI 9128-ENV; Rn; Si; Si detector; airborne particulates; alpha particle; beta particle; calibration factor; moving filter particulates monitor; Calibration; Design optimization; Filters; Instruments; Monitoring; Radiation detectors; Sampling methods; Silicon; System testing; Velocity measurement;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.2002.801669
  • Filename
    1043599