    A Study of Low-Noise Preamplifier Systems for Use with Room Temperature Mercuric Iodide (Hgi2) X-Ray Detectors

  • Author

    Iwanczyk, J.S. ; Dabrowski, A.J. ; Huth, G.C. ; Del Duca, A. ; Schnepple, W.

  • Author_Institution
    University of Southern California Medical Imaging Science Group 4676 Admiralty Way, Suite 932 Marina Del Rey, CA 90291
  • Volume
    28
  • Issue
    1
  • fYear
    1981
  • Firstpage
    579
  • Lastpage
    582
  • Abstract
    An analysis of different preamplification systems for use with room temperature mercuric iodide x-ray detectors has been performed. Resistor-, drain-, and light-feedback preamplifiers have been studied. Energy resolution values of 295 eV (FWHM) for an Fe-55 source (5.9 keV) and 225 eV (FWHM) for a pulser have been obtained with both the detector and the input FET at room temperature using a pulsed-light feedback preamplifier. Improvement in energy resolution by cooling the input FET using a small Peltier element has been discussed.
  • Keywords
    Circuit noise; Energy resolution; FETs; Noise generators; Preamplifiers; Pulse shaping methods; Spectroscopy; Temperature; X-ray detection; X-ray detectors;