    An automated krypton-85 γ-ray stack monitor

  • Author

    Goles, R.W. ; Brauer, F.P.

  • Author_Institution
    Pacific Northwest Laboratory Richland, Washington 99352
  • Volume
    28
  • Issue
    1
  • fYear
    1981
  • Firstpage
    740
  • Lastpage
    744
  • Abstract
    A Ge(Li) γ-ray detector was used in conjunction with a 6-l pressurized (60 psia) well spectroscopy gas cell to selectively detect 85Kr stack emissions. This system was calibrated so as to relate the 514-keV γ-ray counting rate to 85Kr concentration. Counting rate, or concentration, was continuously recorded with a count-rate meter/strip-chart recorder combination and was also time averaged over 15-min intervals using a programmable multichannel analyzer system with cassette readout. Being completely automated, this γ-analysis system required little more than liquid nitrogen service and data record retrieval throughout a 4-month-long sampling period. The sensitivity of this γ-ray analytical system was such as to achieve a minimum detectable 85Kr stack concentration of 2 μCi/m3 for 15-min counting intervals.
  • Keywords
    Atherosclerosis; Calibration; Computerized monitoring; Detectors; Information analysis; Recycling; Spectroscopy; Strips; Temperature; Volume measurement;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1981.4331270