Detection efficiency of Ge(Li) and HPGe detectors for γ-rays up to 10 MeV
Author
Lin, J. ; Henry, E.A. ; Meyer, R.A.
Author_Institution
Lawrence Livermore Laboratory P.O. Box 808 Livermore, CA 94550
Volume
28
Issue
2
fYear
1981
fDate
4/1/1981 12:00:00 AM
Firstpage
1548
Lastpage
1550
Abstract
We have calibrated the relative efficiency up to 9.7 MeV for two coaxial detectors, one Ge(Li) and one high purity Ge. The efficiency curves were determined by using a combination of standard radioactive sources and (n, γ) reactions. Based on the result of this work, the general slope of the two detector efficiency curves appears to be similar and in agreement with earlier work reported by McCallum and Coote. When plotted as a semilogarithmic function of energy the efficiency is linear from 2 to 9.7 MeV.
Keywords
Calibration; Chromium; Coaxial components; Energy capture; Energy resolution; Gamma ray detectors; Gamma rays; Laboratories; Neutrons; Shape measurement;