• DocumentCode
    86048
  • Title

    Incorporating Anatomical Side Information Into PET Reconstruction Using Nonlocal Regularization

  • Author

    Van-Giang Nguyen ; Soo-Jin Lee

  • Author_Institution
    Dept. of Electron. Eng., Paichai Univ., Daejeon, South Korea
  • Volume
    22
  • Issue
    10
  • fYear
    2013
  • fDate
    Oct. 2013
  • Firstpage
    3961
  • Lastpage
    3973
  • Abstract
    With the introduction of combined positron emission tomography (PET)/computed tomography (CT) or PET/magnetic resonance imaging (MRI) scanners, there is an increasing emphasis on reconstructing PET images with the aid of the anatomical side information obtained from X-ray CT or MRI scanners. In this paper, we propose a new approach to incorporating prior anatomical information into PET reconstruction using the nonlocal regularization method. The nonlocal regularizer developed for this application is designed to selectively consider the anatomical information only when it is reliable. As our proposed nonlocal regularization method does not directly use anatomical edges or boundaries which are often used in conventional methods, it is not only free from additional processes to extract anatomical boundaries or segmented regions, but also more robust to the signal mismatch problem that is caused by the indirect relationship between the PET image and the anatomical image. We perform simulations with digital phantoms. According to our experimental results, compared to the conventional method based on the traditional local regularization method, our nonlocal regularization method performs well even with the imperfect prior anatomical information or in the presence of signal mismatch between the PET image and the anatomical image.
  • Keywords
    biomedical MRI; computerised tomography; medical image processing; positron emission tomography; MRI scanner; PET image reconstruction; PET-magnetic resonance imaging scanner; X-ray CT; anatomical boundaries; anatomical edges; anatomical side information; digital phantoms; nonlocal regularization method; positron emission tomography-computed tomography; segmented regions; signal mismatch problem; Positron emission tomography (PET); anatomical priors; nonlocal regularization; penalized likelihood; statistical image reconstruction; Algorithms; Computer Simulation; Image Processing, Computer-Assisted; Magnetic Resonance Imaging; Phantoms, Imaging; Positron-Emission Tomography; Tomography, X-Ray Computed;
  • fLanguage
    English
  • Journal_Title
    Image Processing, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    1057-7149
  • Type

    jour

  • DOI
    10.1109/TIP.2013.2265881
  • Filename
    6522880