Title :
Picoammeter for Ion Chamber Spans Six Decades without Range Switching
Author :
Burkhardt, W. ; Iacopini, M. ; Maranesi, P.
Author_Institution :
Politecnico di Milano, Istituto di Ingegneria nucleare, via Ponzio 34/3 20133 Milano, Italy
Abstract :
Precise current measurements in the range 10-13A to 10-7A can be done without range switching using an active integrator built around a varactor-bridge Op-Amp. The information is supplied by the time required to cross the voltage gap between two thresholds and is carried out by a pulse-width analog signal. Conversion into current or gamma dose rate is done by a simple digital encoder. Suitable contrivances have been adopted in the circuit to minimize leakage and drift effects and to prevent the need of any recalibration of the electrometer.
Keywords :
Capacitance; Capacitors; Circuits; Current measurement; Fluctuations; Leakage current; Operational amplifiers; Q measurement; Switches; Time measurement;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1983.4332276