DocumentCode :
863261
Title :
Improved Energy Resolution Capability of an Imaging Proportional Counter Using Electron Counting Techniques
Author :
Siegmund, O.H.W. ; Clothier, S. ; Culhane, J.L. ; Mason, I.M.
Author_Institution :
Mullard Space Science Laboratory, University College London, Holmbury St Mary, Dorking, Surrey RH5 6NT, UK
Volume :
30
Issue :
1
fYear :
1983
Firstpage :
350
Lastpage :
354
Abstract :
The number of electrons initially created by the absorption of a soft X-ray photon in a parallel plate proportional counter may be counted by detection of the fast light pulses emitted from the resulting electron avalanches. We show that good energy resolution may be achieved (~34% at C-K, 0.28 keV) with this technique. Using the light signal it is also possible to image events by means of an image intensifier, and position resolution of 350 ¿m FWHM at 0.28 keV has been achieved without event centroiding.
Keywords :
Counting circuits; Electromagnetic wave absorption; Electron emission; Energy resolution; Fast light; Image intensifiers; Optical imaging; X-ray detection; X-ray detectors; X-ray imaging;