Title :
Temperature Dependence of NaI(Tl) Decay Constant
Author :
Schweitzer, J.S. ; Ziehl, W.
Author_Institution :
Schlumberger-Doll Research, Ridgefield, Ct.
Abstract :
NaI(Tl) detectors are used for a wide range of applications in different temperature environments. Optimizing NaI(Tl) spectrometer performance under these conditions requires a knowledge of the change in NaI(Tl) decay time constant over the temperature range of spectrometer operation. Measurements of the decay time constant are presented between -25°C and +180°C. Above +25°C the decay time constant decreases slowly, but increases rapidly below +25°C at a rate of ~5 nsec/°C.
Keywords :
Anodes; Detectors; Photomultipliers; Pulse measurements; Pulse shaping methods; Spectroscopy; Temperature dependence; Temperature distribution; Time measurement; Voltage;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :