    A Sealed High Pressure Xenon Filled Imaging Proportional Counter with a Sensitive Area 30cm Square

  • Author

    Thomas, H.D. ; Turner, M.J.L.

  • Author_Institution
    X-Ray Astronomy Physics Department Leicester University Leicester England
  • Volume
    30
  • Issue
    1
  • fYear
    1983
  • Firstpage
    499
  • Lastpage
    502
  • Abstract
    The performance of an imaging proportional counter (IPC) 30cm square is evaluated in this paper. The scaling effects involved in the construction of such a device were investigated using an electrode array alone and a xenon-methane filled IPC. In the IPC, a spatial resolution of 1.2mm FWHM was achieved; and as expected this limitation is shown not to be due to the electronic signal to noise ratio, but to the widely dispersed secondary avalanches initiated by ultra violet photons emitted by excited xenon atoms.
  • Keywords
    Cathodes; Counting circuits; Detectors; Electrodes; High-resolution imaging; Optical imaging; Sensor arrays; Spatial resolution; X-ray imaging; Xenon;