DocumentCode
863883
Title
The High-Density Avalanche Chamber for Positron Emission Tomography
Author
Jeavons, A. ; Hood, K. ; Herlin, G. ; Parkman, C. ; Townsend, D. ; Magnanini, R. ; Frey, P. ; Donath, A.
Author_Institution
CERN, Geneva, Switzerland
Volume
30
Issue
1
fYear
1983
Firstpage
640
Lastpage
645
Abstract
The development of wire chambers for positron emission tomography continues at CERN. In its present form, the basic detector is called a HIgh Density Avalanche Chamber (HIDAC). Owing to the Penning effect, electron avalanche multiplication is obtained in the gamma-ray converter: coincidence time resolution is reduced to 20 ns and stable chamber operation is achieved with a safe, non-polymerizing gas mixture. Spatial resolution remains at 2-3 mm FWHM. A rotating camera consisting of two 20 Ã 20 cm chambers has now been under evaluation at Geneva Hospital for one year. Multilayer printed-circuit techniques are now used to construct chambers with multiple converters, thus raising detection efficiency from 7.5 to 20%. Read-out electronics and back-projecting memory are being developed to handle the high data rate from these chambers. A four-chamber positron camera designed to achieve 100,000 coincidences s-1 is nearing completion.
Keywords
Cameras; Charge carrier processes; Costs; Electrons; Gamma ray detectors; Hospitals; Ionization; Positron emission tomography; Spatial resolution; Wire;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1983.4332347
Filename
4332347
Link To Document