• DocumentCode
    864023
  • Title

    Effect of the Software Coincidence Timing Window in Time-of-Flight Assisted Positron Emission Tomography

  • Author

    Yamamoto, M. ; Ficke, D.C. ; Ter-Pogossian, M.M.

  • Author_Institution
    National Institute of Radiological Sciences 4-9-1 Chiba 260, Japan
  • Volume
    30
  • Issue
    1
  • fYear
    1983
  • Firstpage
    711
  • Lastpage
    714
  • Abstract
    The effect of the software coincidence timing window (SCW) has been studied and demonstrated experimentally using Super PETT I, a time-of-flight (TOF) assisted positron emission tomograph. This method utilizes the TOF information to eliminate unwanted random and scattered coincidence events outside of a region of interest. The effect is substantial for high count rate studies, and can dramatically improve the noise contribution from measured attenuation factors. The effective coincidence window is reduced to 3.3 nsec for reconstructions of the field of view, and reduced to only 2 nsec for measurement of the attenuation factors.
  • Keywords
    Attenuation measurement; Detectors; Image quality; Image reconstruction; Imaging phantoms; Noise measurement; Positron emission tomography; Radioactive decay; Radiology; Timing;
  • fLanguage
    English
  • Journal_Title
    Nuclear Science, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9499
  • Type

    jour

  • DOI
    10.1109/TNS.1983.4332362
  • Filename
    4332362