DocumentCode
864077
Title
Preliminary Results with TOFPET
Author
Mullani, Nizar A. ; Wong, Wai-Hoi ; Hartz, Ross ; Yerian, Keith ; Philippe, Edouard A. ; Gaeta, Jesus M. ; Gould, K.Lance
Author_Institution
University of Texas Medical School at Houston
Volume
30
Issue
1
fYear
1983
Firstpage
739
Lastpage
743
Abstract
TOFPET is a high resolution whole body positron camera which uses time-of-flight information to (a) improve the signal to noise ratio of a reconstructed image and (b) improve the sensitivity of the central slices by using all the interplane coincidences. The first phase of the construction of the camera has been completed, with all the detectors operational and data collected in the list format with a VAX 11/750 computer. Some of the preliminary results obtained with TOFPET such as resolution and sensitivity, as well as, images of Rubidium-82 in a dog heart are presented. The second phase of the construction, which involves the installation of the on-line microprocessors for faster image reconstruction, is being implemented at this time and is expected to be completed by early 1983, at which time the TOFPET camera will be used for routine clinical studies.
Keywords
Cameras; Detectors; Heart; Image reconstruction; Image resolution; Microprocessors; Phase detection; Positrons; Signal resolution; Signal to noise ratio;