Abstract :
We have developed an instrument to measure neutron multiplication by statistical analysis of the timing of neutrons emitted from fissionable material. This instrument is capable of repeated analysis of the same recorded data with selected algorithms, graphical displays showing statistical properties of the data, and preservation of raw data on disk for future comparisons. In our measurements we have made a comparison of the covariance to mean and Feynman variance to mean analysis algorithms to show that the covariance avoids a bias term and measures directly the effect due to the presence of neutron chains. A spherical assembly of enriched uranium shells and acrylic resin reflector/moderator components used for the measurements is described. Preliminary experimental results of the Feynman variance to mean measurements show the expected correlation with assembly multiplication.
Keywords :
Algorithm design and analysis; Assembly; Detectors; Displays; Event detection; Instruments; Laboratories; Neutrons; Statistical analysis; Timing;