Experience with a Double-Compensating Beam Calorimeter
Author
Jarmie, Nelson ; Brown, Ronald E. ; Hardekopf, R.A. ; Martinez, R.
Author_Institution
Los Alamos National Laboratory Los Alamos, New Mexico 87545
Volume
30
Issue
2
fYear
1983
fDate
4/1/1983 12:00:00 AM
Firstpage
1508
Lastpage
1510
Abstract
In an experiment to measure the D(t,¿)n cross section at beam energies of 10 to 120 keV, we have developed a double-compensating beam calorimeter, based on a Swiss Design to measure the particle beam intensity. A Faraday cup is not useful because of considerable charge exchange in the target gas at such low beam energies. We calibrated the calorimeter both with 10- and 3-MeV protons (comparing with a Faraday-cup measurement of the beam flux) and with the heat generated in a precision resistor. Both methods agree and give a calibration accurate to ±0.08% over a range of 10 to 800 mW beam power. Beam powers as low as 5 mW may be used, but with less accuracy. The beam energy must be known in order to calculate the particle intensity. Some difficulties with and peculiarities of the device are discussed.
Keywords
Calibration; Copper; Energy measurement; Particle beam measurements; Particle beams; Particle measurements; Silver; Temperature; Thermistors; Voltage;