Abstract :
An analysis of errors associated with the digital measurement of phase angle between two signals, one of which may be distorted by a harmonic, is presented. All the results were found by running a simulation program on a VAX 11/780 computer. The results are very useful for the users of the phase meters. This technique may introduce large errors for some particular types of input waves. The main purpose of this work is to explain how large the error could be under certain conditions on the input waves
Keywords :
DEC computers; computer aided analysis; digital instrumentation; digital simulation; electric distortion; error analysis; harmonics; measurement errors; phase measurement; phase meters; VAX 11/780 computer; digital phase measurement; distorted waves; harmonic; input waves; measurement error analysis; phase angle; phase meters; simulation program; Computational modeling; Computer errors; Distortion measurement; Error analysis; Goniometers; Harmonic analysis; Harmonic distortion; Phase distortion; Phase measurement; Signal analysis;