Title :
Experimental Results of the Dichotomic Sampling in Circular Ring Positron Emission Tomograph
Author :
Cho, Z.H. ; Hilal, S.K. ; Ra, J.B. ; Hong, K.S. ; Lee, H.S. ; Bigler, R.E.
fDate :
6/1/1983 12:00:00 AM
Abstract :
Experimental verification of the new dichotomic sampling scheme was attempted and applied to a circular ring PET system to improve the sampling thereby the overall system resolution and its performances. In the experiment, two different types of aperture collimators were adopted for high resolution (HR) and very high resolution (VHR) imaging. In HR mode, a resolution of 6.5 mm FWHM was obtained without appreciable degradation in overall sensitivity which represents a threefold resolution improvement over the original system. In phantom studies with HR mode a sensitivity of 4500 counts/sec/¿Ci/cc was obtained for a 20 cm diameter uniform phantom filled with water. VHR mode experiment was also conducted to observe the ultimate resolution capability of DICHOTOM-I system and resolution of 4.2 mm FWHM was obtained at the expense of sensitivity reduction by a factor of four from HR mode experiment.
Keywords :
Apertures; Collimators; Crystals; Degradation; Detectors; High-resolution imaging; Image resolution; Positron emission tomography; Radioactive decay; Sampling methods;
Journal_Title :
Nuclear Science, IEEE Transactions on
DOI :
10.1109/TNS.1983.4332667