• DocumentCode
    86730
  • Title

    Reducing Risks in Type 1 Diabetes Using {cal H}_\\infty Control

  • Author

    Colmegna, Patricio ; Sanchez Peña, Ricardo S. ; Gondhalekar, Ravi ; Dassau, Eyal ; Doyle, Francis J.

  • Author_Institution
    Centro de Sist. y Control, Inst. Tecnol. de Buenos Aires, Buenos Aires, Argentina
  • Volume
    61
  • Issue
    12
  • fYear
    2014
  • fDate
    Dec. 2014
  • Firstpage
    2939
  • Lastpage
    2947
  • Abstract
    A control scheme was designed in order to reduce the risks of hyperglycemia and hypoglycemia in type 1 diabetes mellitus (T1DM). This structure is composed of three main components: an H robust controller, an insulin feedback loop (IFL), and a safety mechanism (SM). A control-relevant model that is employed to design the robust controller is identified. The identification procedure is based on the distribution version of the UVA/Padova metabolic simulator using the simulation adult cohort. The SM prevents dangerous scenarios by acting upon a prediction of future glucose levels, and the IFL modifies the loop gain in order to reduce postprandial hypoglycemia risks. The procedure is tested on the complete in silico adult cohort of the UVA/Padova metabolic simulator, which has been accepted by the Food and Drug Administration (FDA) in lieu of animal trials.
  • Keywords
    diseases; feedback; medical control systems; patient diagnosis; H robust controller; UVA-Padova metabolic simulator; glucose level prediction; hyperglycemia risk reduction; insulin feedback loop; postprandial hypoglycemia risk reduction; safety mechanism; type 1 diabetes mellitus; Diabetes; H infinity control; Insulin; Prediction algorithms; Robust control; H??? control; Type??1 diabetes; hyperglycemia; hypoglycemia;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2014.2336772
  • Filename
    6851161