• DocumentCode
    871916
  • Title

    Temporal Resolution of the Skin Impedance Measurement in Frequency-Domain Method

  • Author

    Fukumoto, Takenori ; Eom, Gwang-Moon ; Ohba, Shigeo ; Futami, Ryoko ; Hoshimiya, Nozomu

  • Author_Institution
    Healthcare Bus. Co., Matsushita Electr., Yokohama
  • Volume
    54
  • Issue
    1
  • fYear
    2007
  • Firstpage
    170
  • Lastpage
    173
  • Abstract
    The temporal-resolution of the frequency-domain method for the identification of the impedance locus depends on the basis frequency used in the current waveform construction, i.e., the higher basis frequency provides the better temporal resolution. The impedance locus can be characterized by the impedance parameters Z0, beta, and taum. The frequency distribution of limited number of data in the impedance locus would affect the accuracy of the estimated impedance parameters. Therefore, we investigated the relationship between the estimation accuracy of the impedance parameters and the frequency coverage of the impedance locus in relatively low to high impedance conditions (dc impedance Z0=51 kOmega-45 MOmega). As the basis frequency, 100 Hz was enough for the usual impedance with Z0 less than 203 kOmega. On the other hand, 10 Hz and 1 Hz were required for the medium-level (Z0=517 kOmega), and high-level (Z0=45 MOmega) impedance, respectively. The required basis frequency, accordingly the temporal resolution, depended much on the central relaxation time taum which affects the frequency distribution on the impedance locus. The results of this study are expected to serve as the reference of the frequency selection in the frequency-domain analysis of the skin impedance
  • Keywords
    bioelectric phenomena; electric impedance; frequency-domain analysis; parameter estimation; skin; 100 Hz; 51 kohm to 45 Mohm; basis frequency; central relaxation time; current waveform construction; frequency-domain analysis; parameter estimation; skin impedance measurement; temporal resolution; Biomedical engineering; Biomedical measurements; Frequency conversion; Frequency domain analysis; Frequency estimation; Frequency measurement; Impedance measurement; Parameter estimation; Skin; Time domain analysis; Basis frequency; central relaxation time; frequency-domain method; skin impedance; temporal resolution; Algorithms; Computer Simulation; Diagnosis, Computer-Assisted; Electric Impedance; Fourier Analysis; Models, Biological; Plethysmography, Impedance; Reproducibility of Results; Sensitivity and Specificity; Skin Physiology; Time Factors;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2006.883742
  • Filename
    4034033