DocumentCode :
873211
Title :
Characterization of a Fast, Sensitive Gamma Detector Utilizing Microchannel Plate Photomultiplier
Author :
Kruse, Herald W. ; Hocker, Lonnie P.
Author_Institution :
Los Alamos National Laboratory P.O. Box 1663 MS D410 Los Alamos, New Mexico 87545
Volume :
31
Issue :
1
fYear :
1984
Firstpage :
387
Lastpage :
389
Abstract :
Sensitivity of a microchannel plate photomultiplier may be enhanced by coupling to Lucite or quartz with small sacrifice of time resolution. Sensitivity enhancement of a factor of 2.4 was achieved at the expense of an increase in time response from 288 ps FWHM to 330 ps. With such time resolution, amplitude resolution was observed to be 17% at 10 ma of output current and varied inversely as the square root of the current.
Keywords :
Gamma ray detection; Gamma ray detectors; Gamma rays; Linear particle accelerator; Microchannel; Photomultipliers; Pulse shaping methods; Signal resolution; Time factors; Time measurement;
fLanguage :
English
Journal_Title :
Nuclear Science, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9499
Type :
jour
DOI :
10.1109/TNS.1984.4333283
Filename :
4333283
Link To Document :
بازگشت