Recent Developments in Detectors for High Spatial Resolution Positron CT
Author :
Yamashita, Y. ; Uchida, H. ; Yamashita, T. ; Hayashi, T.
Author_Institution :
Hamamatsu Photonics K.K. 1126-1 Ichino-Cho Hamamatsu City, Japan
Volume :
31
Issue :
1
fYear :
1984
Firstpage :
424
Lastpage :
428
Abstract :
Experiments using three types of detectors for high spatial resolution positron CT have been carried out as follows : 1. One to one coupling between a BGO scintillator and a thin rectangular photomultiplier tube (PMT). 2. Combining a small position sensing PMT with two BGO scintillators. 3. Combining a large PMT with two or more BGO scintillators and using other detectors for scintillator identification -- a) small PMT´s, b) silicon photodiodes (SiPD). The time resolution obtained in method 1 was poorer than that of the other methods due to the present status of thin PMT performance - poor cathode sensitivity. In method 2 and 3-a), good time resolution and position sensing were obtained. In method 3-b), the peak to valley ratio of 1.8 in pulse height spectrum was obtained with a cooled BGO scintillator and a SiPD.
Keywords :
Anodes; Cathodes; Computed tomography; Coupling circuits; Detectors; Photonics; Positrons; Pulse measurements; Spatial resolution; Timing;