DocumentCode
873640
Title
Dual Energy Attenuation Correction for Single Photon Emission Computed Tomography (SPECT)
Author
Nalcioglu, O. ; Bor, D.
Author_Institution
Department of Radiological Sciences, University of California, Irvine
Volume
31
Issue
1
fYear
1984
Firstpage
590
Lastpage
593
Abstract
Absorption of gairna rays by matter causes the well kncown attenuation artifacts in SPECT. In this paper, we investigate a technique in which projections acquired at two different energies are used. The projections resulting from the two energies are manipulated to extract the true radioisotope distribution within the slice reducing the attenuation problems. Simulation results are presented for different phantoms investigating the amount of artifacts with and without the correction. It is shown that the dual energy technique can be used in obtaining accurate SPECT images with very little attenuation artifacts.
Keywords
Attenuation measurement; Bones; Distributed computing; Energy measurement; Head; Image reconstruction; Imaging phantoms; Performance evaluation; Single photon emission computed tomography; Thickness measurement;
fLanguage
English
Journal_Title
Nuclear Science, IEEE Transactions on
Publisher
ieee
ISSN
0018-9499
Type
jour
DOI
10.1109/TNS.1984.4333325
Filename