Improved Analysis of Implicit RMS Detectors
Author
Abuelma´atti, Muhammad Taher
Author_Institution
King Fahd Univ. of Pet. & Miner., Dhahran
Volume
58
Issue
3
fYear
2009
fDate
3/1/2009 12:00:00 AM
Firstpage
502
Lastpage
505
Abstract
In this paper, the output voltage of the implicit root-mean-square (RMS) detector, which is composed of two voltage-to-current converters, a multiplier/divider, and a low-pass filter and excited by an input voltage comprising a single sinusoid, is approximated using a Fourier cosine series. Using this approximation, closed-form analytical expressions for the output voltage components of the implicit RMS detector are derived. The results obtained are compared with previously published results. This comparison shows that, using the proposed approximation, an improved estimation of the RMS value and the amplitudes of the second- and fourth-harmonic components of the implicit RMS detector can be achieved for any value of the input frequency and is not restricted to large values compared with the cutoff frequency of the low-pass filter.
Keywords
Fourier series; approximation theory; electric sensing devices; harmonic analysis; low-pass filters; multiplying circuits; power convertors; voltage dividers; Fourier cosine series approximation; closed-form analytical expressions; harmonic components; implicit RMS detectors; low-pass filter; multiplier/divider; root-mean-square detector; voltage-to-current converters; Fourier series; harmonics; root-mean-square (RMS) detectors;
fLanguage
English
Journal_Title
Instrumentation and Measurement, IEEE Transactions on
Publisher
ieee
ISSN
0018-9456
Type
jour
DOI
10.1109/TIM.2008.2005262
Filename
4633676
Link To Document