DocumentCode :
873908
Title :
Development of a Portable Microdosimetric Radiation Protection Monitor Covering a Dynamic Range of 120 dB above Noise
Author :
Arbel, A. ; Booz, J. ; Mhuller, K.D. ; Neuhaub, H.
Author_Institution :
KFA Jiulich GmbH, on leave of absence from the Israel Institute of Technology, Haifa, Israel
Volume :
31
Issue :
1
fYear :
1984
Firstpage :
691
Lastpage :
696
Abstract :
A portable microdosimeter has been designed, covering a dynamic range of 120 dB in a single signal processing channel. A transistor is employed as non-linear element and deviation at low and high energies from a truly logarithmic characteristic is corrected by a microcomputer. A description of the basic design and experimental result are given.
Keywords :
Biomedical signal processing; Dynamic range; Employment; Instruments; Microcomputers; Preamplifiers; Protection; Radiation detectors; Radiation monitoring; Signal design;
fLanguage :
English