DocumentCode :
875573
Title :
Microelectrode array for chronic deep-brain microstimulation and recording
Author :
McCreery, Douglas ; Lossinsky, Albert ; Pikov, Victor ; Liu, Xindong
Author_Institution :
Neural Eng. Program, Huntington Med. Res. Inst.s, Pasadena, CA, USA
Volume :
53
Issue :
4
fYear :
2006
fDate :
4/1/2006 12:00:00 AM
Firstpage :
726
Lastpage :
737
Abstract :
We have developed an array of microelectrodes that is suitable for long-term implantation into the subthalamic nucleus (STN) or the globus pallidus and is able to record from single neurons, as well as deliver localized microstimulation. This device can be used to investigate the mechanisms by which deep brain stimulation can ameliorate the symptoms of Parkinson´s disease and other movement disorders, and also may be the basis for a new clinical tool for the treatment of Parkinson´s disease, by capitalizing on the high spatial specificity of intranuclear microstimulation. The array includes 16 activated iridium microelectrodes, 5-6 mm in length, within a cluster approximately 1.8 mm in diameter. We have fabricated the array using materials carrying the USP Category VI classification, and we have developed an apparatus and a procedure for implanting the microelectrode arrays into the deep brain. Ten arrays have been implanted into the STN of domestic cats, and one into the internal segment of the globus pallidus, for 140-415 days. During that time, we were able to record action potentials from individual neurons, on 4 to 8 of the 16 channels. The microelectrode´ active surface areas ranged from 500 to 2 000 μm2. Controlled-current pulses, 26.5 μA in amplitude and 150 μs/phase in duration (4 nC/phase) were used to excite neurons in the cat´s STN. In addition to direct activation, the stimulus modulated the neuronal activity over a distance of at least 1.2 mm from the site of stimulation. These parameters did not induce histologically detectable changes around the tip sites after 35 hours of stimulation at 100 Hz (7 hours of stimulation per day, on 5 successive days), if the electrode´ active surface area was 1 000 μm2 or greater.
Keywords :
arrays; bioelectric potentials; biomedical electrodes; brain; diseases; microelectrodes; neurophysiology; patient treatment; prosthetics; 100 GHz; 140 to 415 day; 26.5 muA; 35 hour; 5 day; 5 to 6 mm; 7 hour; Parkinson disease; action potentials; chronic deep-brain microstimulation; deep brain stimulation; domestic cats; globus pallidus; intranuclear microstimulation; localized microstimulation; long-term implantation; microelectrode array; neurons; subthalamic nucleus; Animals; Brain stimulation; Cats; Electrical stimulation; Microelectrodes; Neural engineering; Neurons; Parkinson´s disease; Satellite broadcasting; Surface treatment; Cats; chronic implant; deep brain stimulation; microstimulation; safety of electrical stimulation; subthalamic nucleus; Action Potentials; Animals; Brain; Cats; Deep Brain Stimulation; Electrodes, Implanted; Equipment Design; Equipment Failure Analysis; Microelectrodes; Monitoring, Physiologic; Neurons; Reproducibility of Results; Sensitivity and Specificity;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2006.870215
Filename :